PMS PROCEDURE (continued)
PM Certification Procedure for Fabius Tiro Anesthesia System
1. Use the PM Certification form P/N 4118460 for the Fabius Tiro Anesthesia System.
2. Completely fill in the header information.
3. Perform the vapor concentration test on all Dräger vapor vaporizers every six
months in accordance with SP00073 at a six month maximum interval. Perform the
vaporizer concentration test on all Desflurane vaporizers in accordance with
SP00091 for fixed mount vaporizers and SP00189 for user removable D-tec
vaporizers at a six month maximum interval. For every vaporizer tested, fill out a
"VAPOR VAPORIZER CALIBRATION CHECK" label (part # S010016). Information
on this label shall include your signature, type of agent, date tested, test results @
1%, 2.5%, 4% for H, E, I, or S vaporizers, or @ 4%, 6%, 10%, 12%, 16% for
Desflurane vaporizers, and a PASS or FAIL indication. This label shall be attached
to the upper right side of the vaporizer. If vaporizer fails the concentration
verification, internal leak, or exclusion system tests, check "NO" in the
"RECOMMENDED FOR USE" section on the PM Certification form.
Place a "CAUTION DO NOT USE" label (part # 4114327) on the vaporizer, and
issue a departmental alert. The TSR shall also seek permission from the equipment
operator to remove the failed vaporizer from the machine and install a replacement
vaporizer or an adapter block onto the mount.All nonfunctional Dräger vaporizers
must be removed from service for machine to receive certification.
4. Proceed with PM Certification in accordance with Section 8.0. If any tests fail refer
to the "Failure Codes" listing in NAD Recommendations Guidelines Index, P/N
S010250, to determine correct certification level starting point. Failure codes shall
be documented on the "RECOMMENDATIONS / GENERAL COMMENTS" section
of the PM Certification form and on the Executive Summary if applicable. If a test
fails that has not been identified by the "Failure Codes" list, consult with Draeger
Medical, Inc. to assess the proper certification level.