Advertisement

Operating Concept

Ventilation Controls

Ventilation Controls

1 Keys for selecting the ventilation modes:
– IPPV
– SIMV
– BIPAP
and
– other modes
2 Keys for selecting / setting ventilation parameters:
– Tidal volume VT
– Inspiration time Tinsp
– Frequency f
– Inspiratory flow Flow
– Inspiratory pressure Pinsp
– Pressure-assisted spontaneous breathing ∆PASB
– Positive end-expiratory pressure PEEP
– Pressure rise time Ramp
– O
concentration O
2
2
– Sensitivity Trigger
3 Central "turn-and-push" rotary knob for setting the
parameters:
To set = turn the rotary knob
To confirm setting = press the rotary knob.

Setting ventilation parameters

2 To set a ventilation parameter = press the
corresponding parameter key.
The yellow LED in the key lights up.
3 To set the value of the ventilation parameter =
turn the rotary knob. The value is displayed next to the
parameter key.
3 To confirm the value = press the rotary knob.
The yellow LED goes out.

Selecting the ventilation mode

1 Hold down the appropriate key for about 3 seconds
or
press the appropriate key briefly and
confirm = press the rotary knob.
The selected ventilation mode will now be activated.
For detailed instructions on setting the ventilation modes,
see page 25.
14
D
2
Evita 2 dura
3
1