Software Update Procedure
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
8. Set the machine power switch to ON. The bootstrap load of the FLASH
programmer load program then proceed according to the status mes-
sages on the PC. The following screen will be displayed:
Figure 4
Image File Transfer Execution Screen
9. Follow the same sequence as a normal download procedure to access
the software. Refer to the appropriate Service Procedure. Once the load
program is executing on the machine, the image load will proceed
according to messages on the machine and the PC until a message on
the machine indicates completion.
10. Close the PC Loader screen and power down the PC.
11. Disconnect the boot strap download cable.
12. Return the boot selector jumper at JP2 (on Control PCB P/N 4116632) or
JP3 (on Control PCB P/N 4118079) to the Boot Flash position.
Figure 5
Boot Flash Position