Function Description
When the machine is operating in the Manual mode, the pump on the pneu-
matic assembly (and the ventilator) is stopped, and the spring-loaded APL
bypass valve in the breathing system closes, directing exhaled gas through
the APL valve.
42
Dräger Medical AG & Co. KGaA
6020.002
Revision B Released