Ventilation Controls; Setting Ventilation Parameters; Selecting The Ventilation Mode - Dräger Medical Savina Operating Instructions Manual

Intensive care ventilator
Hide thumbs Also See for Savina:
Table of Contents

Advertisement

Operating Concept

Ventilation Controls

Setting Ventilation Parameters

Selecting the Ventilation Mode

Operating Concept
Ventilation Controls
1 Keys for selecting the ventilation modes:
–CMV A/C
–SIMV
–CPAP/PS
–PCV+
2 Keys for selecting / setting ventilation parameters:
– Tidal volume V
T
– Inspiration time Tinsp
– Frequency Rate
– O
-concentration O
2
2
– Inspiratory pressure Pinsp
– Pressure support ∆P
supp.
– Positive End-Expiratory Pressure PEEP
3 Central "turn and push" rotary dial knob for setting
parameters and for selecting screen parameters.
To set, turn dial knob
To confirm setting, press dial knob.
Setting Ventilation Parameters
2 To select a ventilation parameter, press the respective
parameter key.
The yellow LED in the key lights up.
3 To set value of a ventilation parameter, turn dial knob.
The value is displayed next to the parameter key.
3 To confirm value, press dial knob.
The yellow LED in the key goes out.
NOTE: A new parameter value only becomes effective upon
confirmation.
Derived parameters are displayed in an information window
on the main screen page during the setting procedure.
Selecting the Ventilation Mode
1 Hold down the appropriate key for about 3 seconds
or
press key briefly and
3 confirm by pressing dial knob
The selected ventilation mode will now be activated.
For detailed instructions on setting ventilation modes,
see page 52 and following.