Guidance And Manufacturer's Declaration-Electromagnetic Immunity - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
4
Guidance and manu-
facturer's declara-
tion-electromagnetic
immunity
Dräger Medical Systems, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Electromagnetic Testing and Compliance
Immunity test
IEC 60601 test
Voltage dips,
<5% Ut
short interrup-
(>95% dip in Ut)
tions, and volt-
for 0.5 cycle
age variations on
power supply
40% Ut
input lines
(60% dip in Ut)
IEC 61000-4-11
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% VT
(>95% dip in Ut)
for 5 seconds
Power fre-
3 A/m
quency
(50/60 Hz) mag-
netic field
IEC61000-4-8
NOTE
Ut is the a.c. mains voltage prior to application of the test level.
The Fabius Tiro is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the Fabius Tiro should assure that it
is used in such an environment.
Table 3
Electromagnetic Immunity
Immunity
IEC 60601
test
test level
6020.002
Compliance
level
level
Complies
Complies
Compliance
Electromagnetic environ-
level
Portable and mobile RF com-
munications equipment
should be used no closer to
any part of the Fabius Tiro,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the fre-
quency of the transmitter.
Revision B Released
Recommendations
Mains power quality
should be that of a
typical commercial or
hospital environ-
ment. The Fabius
Tiro provides battery
back-up in the event
of a power failure.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environ-
ment.
ment-guidance
55