Replaceable Parts

Assembly Replacement Sequence

Assembly Replacement Sequence
The following steps outline the sequence to be performed to replace an assembly in the
network analyzer.
Step 1. Identify the faulty group. Begin with
with the appropriate troubleshooting chapter that identifies the faulty assembly.
Step 2. Order a replacement assembly. Refer to this chapter.
Step 3. Replace the faulty assembly and determine what adjustments are necessary.
Refer to
Chapter 7, "Repair and Replacement Procedures."
Step 4. Perform the necessary adjustments. Refer to
Step 5. Perform the necessary performance tests. Refer to
Adjustments."
6-4
PNA Series Microwave Network Analyzers
Chapter 4, "Troubleshooting."
Chapter 3, "Tests and Adjustments."
Chapter 3, "Tests and
Service Guide E8364-90001
E8362A, E8363A, E8364A