Fitting The New Source Enclosure And Probe Adjuster Assembly Probe Seals - Waters Vion IMS QTof Overview And Maintenance Manual

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Figure 5–100: Removing the source enclosure, nebulizer gas, and desolvation gas
seals
Desolvation gas seal
Source enclosure seal
Nebulizer gas seal
Warning:
corrosive materials, dispose of all waste materials according to local environmental
regulations.
4. Dispose of all the seals in accordance with local environmental regulations.
5.24.2 Fitting the new source enclosure and probe adjuster assembly
probe seals
Required materials
• Chemical-resistant, powder-free gloves
• Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
• New seals
• Lint-free cloth
To avoid spreading contamination with biologically hazardous, toxic, and
March 17, 2021, 715005046 Ver. 03
Page 145