Fitting The New Source Enclosure And Probe Adjuster Assembly Probe Seals - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

4.
Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure and probe adjuster assembly
probe seals
Required materials
Chemical-resistant, powder-free gloves
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
New seals
To fit the new source enclosure and probe adjuster assembly probe seals:
1.
Ensure that all the grooves for seals are free from dirt and debris.
If contamination is present, use 1:1 methanol/water, applied to a
Tip:
lint-free cloth, to carefully clean the grooves.
Requirement:
correctly located in the groove when fitting them to the source enclosure.
2.
Fit the new source enclosure seal onto the source enclosure.
Start by feeding the seal into the groove at the bottom right-hand
Tip:
corner, and then working around the seal in a counterclockwise
direction.
To avoid spreading contamination, dispose of all
Warning:
the seals according to local environmental regulations. The
seals can be contaminated with biohazards or toxic
materials.
To avoid personal contamination with biohazards
Warning:
or compounds that are toxic or corrosive, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the source components.
Ensure that the tails of the source enclosure seals are
Replacing the source assembly seals
February 28, 2014, 715002212 Rev. C
201