Figure 5–118: Connecting the PEEK female-to-male adapter
Flangeless nut
Flangeless ferrule
PEEK female-to-male adapter
d. Slide the transparent, 1/16-inch, flangeless nut and blue, 1/16-inch, flangeless
ferrule over the tubing.
Rationale:
These components provide fail-safe pressure relief in the event of a
blockage.
e. Insert the tubing in the female-to-male adapter in port 7 and tighten the fittings.
4. Using a long, finger-tight fitting, connect 1/16-inch, 1,000-mm, natural-colored PEEK tubing
to port 1 of the sample selector valve and thread the tubing into the waste port.
Figure 5–119: Connecting the PEEK tubing
Waste port
Figure 5–120: Long, finger-tight fitting
Long, finger-tight fitting