Figure B–4: Pump control and grounding connections
Pump control connection
Grounding connector and grounding screw
Pump control cable (attaches to vacuum hose)
Notes:
• To connect the grounding connector, you must remove the grounding screw and
washer, and refit them with the grounding connector in place.
• Connect the pump control cable to the upper backing pump connector. Leave the lower
connector, marked "Delayed backing pump control", disconnected.