Waters Vion IMS QTof Overview And Maintenance Manual page 95

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

End cover
Rubber gasket
Probe cable
3. Unscrew and remove the nebulizer adjuster knob.
Figure 5–39: Removing the nebulizer adjuster knob
Nebulizer adjuster knob
Probe cable
ESI probe
4. Unscrew and remove the ESI probe tip by holding the probe shaft steady using the 7-mm
wrench and unscrewing the probe tip using the 10-mm wrench, as shown in the following
figure:
March 17, 2021, 715005046 Ver. 03
Page 95