4 Maintenance Procedures
4.
Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure and probe adjuster assembly
seals
Required materials
Chemical-resistant, powder-free gloves
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
New seals
176
March 14, 2014, 715004387 Rev. B
Nebulizer gas seal
TP03164
To avoid spreading contamination, dispose of the
Warning:
seals in accordance with local environmental regulations.
The seals can be contaminated with biohazardous or toxic
materials.
Desolvation gas seal
Source enclosure seal