To remove the StepWave assembly from the source assembly:
Notice:
To avoid damaging the StepWave ion guide assembly when removing it from
the source assembly, use only these tools:
• Seal-breaker and locator tool
• StepWave assembly removal and insertion tool
Notice:
To avoid damaging the StepWave ion guide assembly, handle it and its
components carefully throughout the cleaning procedure.
1. Position the seal breaker and locator tool so that it engages as follows:
• Pin with the hole on the adapter housing directly under the end of the StepWave ion
guide assembly.
• Lip behind the StepWave ion guide assembly's ion guide cap.
Figure 5–29: Seal breaker and locator tool
Lip
Pin
Handle
March 17, 2021, 715005046 Ver. 03