Figure 5–35: Removing the nickel gasket
Nickel gasket
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
4. Dispose of the nickel gasket in accordance with local environmental regulations.
5. If the probe tip is damaged, dispose of it in accordance with local environmental
regulations, and continue this procedure with a new ESI probe tip.
6. Insert the new nickel gasket into the ESI probe tip so that the gasket is seated around the
narrow, stainless steel tube inside the probe tip.
Figure 5–36: Inserting the new nickel gasket
Nickel gasket
ESI probe tip
Stainless steel tube
7. Carefully slide the probe tip onto the ESI probe, ensuring that the capillary feeds through
the stainless steel tube inside the probe tip.
March 17, 2021, 715005046 Ver. 03
Page 92