Waters Vion IMS QTof Overview And Maintenance Manual page 163

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Sample selector valve
Divert valve
Waste
To reference sprayer
Waste
All tubing is PEEK, except for that comprising the connection between the pump and sample
selector valve, which is stainless steel.
Table 5–5: Tubing dimensions (sample delivery system)
Valve/port
Connection
Sample/1
Waste
reservoir
Sample/2
Wash reservoir 0.020
Sample/3
Diverter valve
Sample/4
Reservoir
Bottle A
Sample/5
Reservoir
Bottle B
Sample/6
Reservoir
Bottle C
Sample/7
Analyte pump
Diverter/4
Waste
Required materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• Needle-nose pliers
• The Vion IMS QTof Fluidics Tubing and Fitting Kit
Tip:
This kit contains components for both the sample and NanoSpray system's plumbing.
To plumb the analyte system:
Requirement:
You must finger-tighten all PEEK fittings.
1. Open the access door to the IntelliStart fluidics.
ID (inch)
OD (inch)
0.040
1/16
1/16
0.005
1/16
0.020
1/16
0.020
1/16
0.020
1/16
0.040
1/16
0.040
1/16
March 17, 2021, 715005046 Ver. 03
Page 163
Color
Length (mm)
Natural
1,000
Orange
1,000
Red
200
Orange
680
Orange
680
Orange
680
N/A
500
Natural
1,000