Figure 5–14: Cone extraction tool
Collar
3. Insert the collar in the sample cone.
Figure 5–15: Inserting the cone extraction tool
Insert the collar
Notice:
To avoid damaging the sampling cone, which is fragile, do not place it
on its tip. Always place it on its flanged base.
4. Rotate and lift the tool and collar to remove the sample cone from the cone gas nozzle.
Figure 5–16: Removing the sample cone
Rotate the tool
March 17, 2021, 715005046 Ver. 03