Waters Vion IMS QTof Overview And Maintenance Manual page 157

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Table 5–4: Tubing dimensions (sample delivery system)
Valve/port
Connection
Sample/1
Waste
reservoir
Sample/2
Wash reservoir 0.020
Sample/3
Divert valve
Sample/4
Reservoir
Bottle A
Sample/5
Reservoir
Bottle B
Sample/6
Reservoir
Bottle C
Sample/7
Analyte pump
divert/4
Waste
Required materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• The Vion IMS QTof Fluidics Tubing and Fitting Kit
Tip:
This kit contains components for both the sample and LockSpray system's plumbing.
To plumb the analyte system:
Warning:
To avoid eye injury from the PEEK tubing, use eye protection when
performing this procedure.
1. Open the fluidics access door.
2. Unscrew and remove the thumbscrew securing the fluidics pump access panel and remove
the panel.
3. Using the stainless steel tubing, connect the sample pump to port 7 of the sample selector
valve.
Requirement:
tighten the fittings appropriately.
Recommendation:
chamber to allow the fluidics panel to be replaced later.
ID (inch)
0.040
0.005
0.020
0.020
0.020
0.040
0.040
The steel tubing requires high-pressure seals at each end. You must
Feed the steel tubing through the notch in the side of the fluidics
March 17, 2021, 715005046 Ver. 03
Page 157
OD (inch)
Color
1/16
Natural
1/16
Orange
1/16
Red
1/16
Orange
1/16
Orange
1/16
Orange
1/16
N/A
1/16
Natural
Length (mm)
1,000
1,000
200
680
680
680
500
1,000