Figure 5–77: Removal of the reference probe inlet connector from within the source
housing
Source housing
Reference probe inlet connector
7. Rotate the reference sprayer assembly clockwise through 90 degrees and remove it from
the reference sprayer support assembly.
Tip:
Remove the baffle to improve access to the reference sprayer assembly, if required.
To remove the baffle, use a 2-mm hex wrench.
Figure 5–78: Removal of the reference sprayer assembly
Reference sprayer support assembly
Reference sprayer assembly