Locating notch on sprayer assembly
Reference sprayer assembly stem
5. Rotate the reference sprayer assembly 90 degrees counterclockwise to lock it in place.
Figure 5–84: Locking the reference-sprayer assembly
6. Rotate the reference probe inlet connector until its groove aligns with the source housing
notch.
7. Insert the reference probe inlet connector up to the notch through the hole in the source
housing.
Figure 5–85: Inserting the reference probe into the source housing
Source housing
Source housing access hole
Source housing notch
Bayonet fixing notch in reference probe PEEK inlet connector