4.
Dispose of all the seals in accordance with local environmental
regulations.
Fitting the new source enclosure and probe adjuster assembly
probe seals
Required materials
Chemical-resistant, powder-free gloves
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
New seals
To fit the new source enclosure and probe adjuster assembly probe seals:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Ensure that all the grooves for seals are free from dirt and debris.
If contamination is present, use 1:1 methanol/water, applied to a
Tip:
lint-free cloth, to carefully clean the grooves.
Caution:
correctly located in the groove when fitting them to the source
enclosure.
2.
Fit the new source enclosure seal to the source enclosure.
Start by feeding the seal into the groove at the bottom right-hand
Tip:
corner, and then working around the seal in a counterclockwise
direction.
The seals can be contaminated with biohazardous
Warning:
and/or toxic materials. Dispose of them according to local
environmental regulations.
The source components can be contaminated with
Ensure that the tails of the source enclosure seals are
Replacing the source assembly seals 4-107