Waters Vion IMS QTof Overview And Maintenance Manual page 162

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Important:
When the fluidics system is configured for NanoLockSpray operation, the LC flow
connects directly to the source sample inlet, and the divert valve connects to the reference
sprayer. During NanoLockspray operation, setting the fluidics to LC in the software routes the
reference pump to the source reference sprayer, and setting the fluidics to Infusion in the
software routes the sample pump to the source reference sprayer. The NanoLockSpray
configuration facilitates optimization of the source parameters (see the instrument's online Help
file for information about optimizing the source).
Figure 5–124: NanoLockSpray sample fluidics delivery tubing schematic
Wash
Pump line
Wash
Pump line
March 17, 2021, 715005046 Ver. 03
Page 162