Figure 5–106: Long, finger-tight fitting
Long, finger-tight fitting
4. Using a long finger-tight fitting, connect orange, 1/16-inch, 680-mm, PEEK tubing from port
1 of the LockSpray selector valve, through tubing guide A, and into reservoir bottle A,
ensuring that the tubing reaches the bottom of the bottle.
Figure 5–107: Connecting the PEEK tubing
Reservoir bottle A
Tubing guide A
Tips:
Follow these suggestions if you encounter difficulty threading the tubing through the
guides:
• Thread the tubing upward from the bottom hole.
• Using needle-nose pliers, make a 20° bend, 10 mm from the end of the tubing. Then
rotate the tubing, as necessary, as you thread it through the guide.
5. Using a long finger-tight fitting, connect orange, 1/16-inch, 680-mm, PEEK tubing from port
2 of the LockSpray selector valve through tubing guide B, and into the reference reservoir
bottle in fluidics port B.
Note:
If you are using a 250-mL bottle in fluidics port B, ensure that you pull the tubing far
enough through the tubing guide to allow it to reach the bottom of the bottle.
6. Using a long finger-tight fitting, connect orange, 1/16-inch, 680-mm, PEEK tubing from port
3 of the LockSpray selector valve through tubing guide C, and into reservoir bottle C.
7. Use a long finger-tight fitting to connect orange, 1/16-in, 1,000-mm, PEEK from port 6 of
the LockSpray selector valve to the wash bottle.
March 17, 2021, 715005046 Ver. 03