Waters Vion IMS QTof Overview And Maintenance Manual page 38

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Figure 3–2: ESI probe mounted on the LockSpray source enclosure
Probe locking ring
Source enclosure release
ESI high-voltage cable
Vernier probe adjuster
ESI probe
4. Tighten the probe locking ring to secure the probe in place.
Tip:
An automatic pressure test is performed when the probe is correctly seated in
position.
5. Connect the ESI probe's cable to the high-voltage connector.
6. Open the access door to the fluidics valves (see the figure in
system (Page
24)).
Warning:
steel tubing to connect the divert valve to the probe; use the PEEK tubing
supplied with the instrument.
7. Using a long, finger-tight fitting, connect 0.004-inch ID (or greater) tubing from port 2 (the
top port) of the divert valve to the ESI probe.
8. Secure the union using a PEEK finger-tight nut and ferrule.
To avoid harmless, static-like electric shock, do not use stainless
March 17, 2021, 715005046 Ver. 03
Page 38
Physical layout of the Fluidics