Waters Vion IMS QTof Overview And Maintenance Manual page 154

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Figure 5–110: Long finger-tight PEEK nut and 1/32-inch, PEEK ferrule
Figure 5–111: 1/32-inch, 6-40 Valco compression fitting assembly
12. Using a 1/32-inch, Valco component fitting assembly, connect the red, 1/32-inch, 60-mm,
PEEK tubing between the flow sensor and the grounded union, where you use a use a
short finger-tight nut and 1/32-inch ferrule.
Important:
The narrower, 1/32-inch tubing can slip out of the port. After tightening the
fitting, gently tug the tubing, to ensure it is secure in the port.
Figure 5–112: Connecting the PEEK tubing
Grounded union
Figure 5–113: Short finger-tight nut and 1/32-inch ferrule
Warning:
steel tubing to connect parts; use the PEEK tubing supplied with the instrument.
13. Connect the grounded union to the LockSpray source's reference probe. For the
LockSpray source, use 1/16-inch PEEK tubing connected as follows (for the
NanoLockSpray source, see
NanoLockSpray operation (Page
a. At the grounded union end, use a long finger-tight fitting.
To avoid harmless, static-like electric shock, do not use stainless
Plumbing the fluidics sample delivery system for
161)):
March 17, 2021, 715005046 Ver. 03
Page 154