Groove in reference probe PEEK inlet connector
8. Push the reference probe inlet connector upwards until the bayonet groove is above the
notch.
9. Rotate the reference probe inlet connector until the bayonet groove aligns with the notch
on the source housing.
Figure 5–86: Alignment of notch and groove
Source housing notch
Bayonet groove
10. Push down on the reference probe inlet assembly to lock it into position within the bayonet
fixing.
Figure 5–87: Locking the reference probe inlet assembly into the bayonet fixing
11. Refit the source enclosure (see
Fitting the source enclosure to the instrument (Page
March 17, 2021, 715005046 Ver. 03
Page 132
56)).