Waters Vion IMS QTof Overview And Maintenance Manual page 134

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Figure 5–88: Removing the NanoLockSpray reference probe's fixing screws
Fixing screw
NanoLockSpray reference probe
Warning:
reference-probe tip is an exposed, fused-silica TaperTip emitter that is sharp and
fragile.
7. Remove the NanoLockSpray reference probe from the probe adjuster assembly.
8. Unscrew the TaperTip-emitter PEEK coupler, and remove the TaperTip emitter from the
union.
Figure 5–89: Removing the TaperTip emitter
TaperTip-emitter PEEK coupler
Capillary PEEK coupler
Capillary with PEEK protective sleeve
To avoid puncture wounds, handle the probe with care. The
March 17, 2021, 715005046 Ver. 03
Page 134