Figure 5–40: Removing the ESI probe tip
7-mm wrench
10-mm wrench
Probe tip
Probe shaft
5. Remove the nickel gasket from the probe tip.
Figure 5–41: Removing the nickel gasket
Nickel gasket
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
6. Dispose of the nickel gasket in accordance with local environmental regulations.
Warning:
care.
To avoid puncture wounds, handle sharp parts and materials with
March 17, 2021, 715005046 Ver. 03
Page 96