Removing The Ionsabre Ii Probe - Waters Vion IMS QTof Overview And Maintenance Manual

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

• If you are replacing the tubing supplied with the instrument, you must minimize the
length of the tube connecting the divert valve to the IonSABRE II probe. Doing so
minimizes delays and dispersion.
• When cutting the tubing to length, cut it squarely.
6. Close the access door.
7. Install the corona pin (see

3.3.2 Removing the IonSABRE II probe

Required materials
• Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe:
Warning:
surfaces marked with the high-voltage warning symbol, ensure that the mass
spectrometer is in Standby mode.
Warning:
source; these components can be hot.
1. Prepare the instrument so that you can work safely on the source (see
instrument for working on the source (Page
2. Remove the corona pin (see
3. Disconnect the divert valve tubing from the IonSABRE II probe.
4. Unscrew the probe locking ring.
5. Carefully remove the probe from the probe adjuster assembly.
Installing the corona pin in the source (Page
To avoid harmless, static-like electric shock, before you touch any external
To avoid burn injuries, take great care while working with the probe and
Removing the corona pin from the source (Page
March 17, 2021, 715005046 Ver. 03
Page 43
53)).
57)).