Notice:
To avoid distorting the drip tray or causing the drain cup to leak, restrain
the drain cup when attaching or removing the waste tubing.
2. Slide a drain line over the barbed fitting of the drain cup, located below the source
enclosure.
Figure B–8: Connecting liquid waste lines
Barbed fitting
Source enclosure
Warning:
To prevent leakage of biologically hazardous, toxic, or corrosive materials:
• Do not crimp or bend drain lines. A crimp or bend can impede flow to the waste
container.
• Empty the waste container before the lower ends of the drain tubes are covered by
waste solvent.
3. Slide a drain line over the barbed fitting on the base of the bottle tray on top of the
instrument.
4. Route the waste lines to the waste container. If necessary, shorten the waste tubes so that
their ends are above the surface of the waste solvent.
March 17, 2021, 715005046 Ver. 03
Page 189