Connecting The Contact-Closure Cable To An Acquity Lc - Waters Vion IMS QTof Overview And Maintenance Manual

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Figure B–14: Inserting the connector with the signal cable
Clamp
Cable leads
6. Place the second connection cover over the first cover, and snap it into place.
Figure B–15: Connection cover
Signal connector
Connection cover
B.10 Connecting the contact-closure cable to an ACQUITY
LC
The contact-closure cable connects the mass spectrometer's Gas Fail Interlock signal connectors
to the ACQUITY LC system's Stop Flow In signal connectors.
This connection ensures that flow from the ACQUITY LC is stopped if the nitrogen supply to the
mass spectrometer's source is insufficient.
March 17, 2021, 715005046 Ver. 03
Page 196