To install the ESI probe:
Warning:
To avoid harmless, static-like electric shock, before you touch any external
surfaces marked with the high-voltage warning symbol, ensure that the mass
spectrometer is in Standby mode.
1. Prepare the instrument so that you can work safely on the source (see
instrument for working on the source (Page
Warning:
2. Remove the protective sleeve, if fitted, from the ESI probe tip.
3. With the probe label facing you, carefully slide the ESI probe into the hole in the probe
adjuster assembly, ensuring that the probe location dowel aligns with the location hole in
the probe adjuster assembly.
Figure 3–1: ESI probe insertion into the probe adjuster assembly
Probe label
Probe location dowel
Location hole of the probe adjuster assembly
To avoid puncture wounds, handle the probe with care.
March 17, 2021, 715005046 Ver. 03