Cleaning The Sample Cone And Cone Gas Nozzle - Waters Vion IMS QTof Overview And Maintenance Manual

Hide thumbs Also See for Vion IMS QTof:
Table of Contents

Advertisement

Remove the sample cone
5. Remove the O-ring from the sample cone.
Figure 5–17: O-ring removed from the sample cone
Cone gas nozzle
Sample cone
O-ring
Cone gas nozzle handle
Warning:
corrosive materials, dispose of all waste materials according to local environmental
regulations.
6. If the O-ring shows signs of deterioration or damage, dispose of it in accordance with local
environmental regulations.

5.13.3 Cleaning the sample cone and cone gas nozzle

Required materials
• Chemical-resistant, powder-free gloves
• Appropriately sized glass vessels in which to completely immerse components when
cleaning. Use only glassware not previously cleaned with surfactants.
• HPLC-grade (or better) methanol
• HPLC-grade (or better) water
• Formic acid
• Ultrasonic bath
To avoid spreading contamination with biologically hazardous, toxic, and
March 17, 2021, 715005046 Ver. 03