Figure 5–131: Connecting the PEEK tubing
Divert valve
8. Using a long, finger-tight fitting, connect 1/16-inch, 1,000-mm, natural-colored PEEK tubing
to port 4 of the divert valve and thread it securely into the waste port.
Tip:
This is the same waste port as that used in step 3.
Figure 5–132: Connecting the PEEK tubing
Waste port
9. Connect port 2 of the divert valve to the source probe's reference sprayer.
10. Close the access doors to the IntelliStart Fluidics system.
Important:
doors to the fluidics system.

5.27 Replacing the instrument's fuses

Warning:
supply before replacing fuses. The mass spectrometer has two fuses and uses double
pole/neutral fusing circuitry. Circuits can remain live even when one fuse has blown.
Ensure that the tubing does not become trapped when you close the access
To avoid electric shock, disconnect the mass spectrometer from the power
March 17, 2021, 715005046 Ver. 03
Page 167