Requirement:
Push the tubing to the bottom of the wash solution, and secure the tubing
to prevent it floating to the surface during use.
8. Using a long finger-tight fitting, connect 1/16-inch, 1,000-mm, natural-color PEEK tubing to
port 5 of the LockSpray selector valve, and thread the tubing into the waste port.
Figure 5–108: Connecting the PEEK tubing
Waste port
Tip:
The liquid waste system collects waste without requiring a fitted connection. Waste
drains through the connector at the base of the instrument and into a waste bottle, see
Connecting liquid waste lines (Page
9. If the system lacks an optional flow sensor and grounded union (used for nanoflow
operation), skip to step 14.
10. Using a long finger-tight PEEK nut and 1/32-inch PEEK ferrule, connect the red, 1/32-in,
200-mm, PEEK tubing from port 4 of the LockSpray selector valve to the top of the flow
sensor.
11. Connect the 1/32-in, 6-40 Valco compression fitting assembly.
Figure 5–109: Connecting the Valco compression fitting assembly
Flow sensor
188).