5.24.1 Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
1. Remove the source enclosure from the instrument (see
from the instrument (Page
2. Use the O-ring removal kit (see
remove the probe and nebulizer gas seals from the probe adjuster assembly.
Figure 5–99: Probe and nebulizer gas seals
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
3. Use the O-ring removal kit to carefully remove the source enclosure, nebulizer gas, and
desolvation gas seals from the source enclosure.
54)).
Removing O-rings and seals (Page
March 17, 2021, 715005046 Ver. 03
Page 144
Removing the source enclosure
62)) to carefully