3. Remove the LockSpray II source enclosure from the instrument (see
enclosure from the instrument (Page
4. Wait 10 minutes for the source ion block to cool.
5. Pull up and rotate the reference probe inlet connector to overcome the bayoneted lock,
and to align the indicator arrow on the source housing with the groove on the inlet
connector.
Figure 5–75: Unlocking the inlet connector from the source housing
Inlet connector groove
Source housing indicator arrow
6. Push down on the reference probe inlet connector to release it from the source housing.
Figure 5–76: Releasing the unlocked reference probe from the source housing
Tip:
If the reference probe inlet connector is difficult to remove, rotate the source housing
and remove the inlet connector from within.
Warning:
Warning:
and source; these components can be hot.
54)).
To avoid puncture wounds, handle the probe with care.
To avoid burn injuries, take great care while working with the probe
March 17, 2021, 715005046 Ver. 03
Page 126
Removing the source