10. Dispose of the used reference sprayer probe in accordance with local environmental
regulations.
5.20.2 Installing the new reference probe
Required materials
• Chemical-resistant, powder-free gloves
• 4-mm open-end wrench
• 2-mm hex wrench
To install the new reference probe:
1. Insert the capillary end of the new reference probe into the reference sprayer assembly,
ensuring that it is fully seated.
Figure 5–81: Inserting the reference sprayer capillary into the sprayer assembly
Reference sprayer assembly
Reference sprayer probe capillary
Reference probe
2. Screw the PEEK finger-tight fitting clockwise until you hear an audible click.
Note:
This seats the capillary correctly and avoids dead volume.
March 17, 2021, 715005046 Ver. 03
Page 129