Otto Bock Meridium 1B1-2 Series Instructions For Use Manual page 79

Hide thumbs Also See for Meridium 1B1-2 Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 43
15.2.4 Status signals
Battery charger is connected
LED on
LED on
power sup­
battery
ply
charger
Battery charger disconnected
Beep sig­
Vibration
nal
signal
1 x short
1 x short
3 x short
Battery charge level
Battery
charger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
Battery is fully charged (or connection with product is interrupted).
Take note of the confirmation signal for differentiation.
When the battery charger is connected or disconnected, a self-test is conducted and confirmed by
a beep and vibration signal.
The battery is fully charged if this signal is heard.
If no signal is emitted, the connection to the product is interrupted.
15.3 Directives and manufacturer's declaration
15.3.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (see page 51).
Electromagnetic emissions
Interference measure­
ments
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according
IEC 61000-3-3
1B1-2=* Meridium
Event
Power supply and battery charger operational
Result
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery charger. If the
beep signal sounds again, product maintenance should be carried out by an autho­
rised Ottobock Service Center.
The product can be used without restrictions. However, vibration signals may not be
generated.
Conduct the self-test again by connecting/disconnecting the battery charger. If no
beep/vibration signal is emitted after connecting/disconnecting the battery charger
again, the product must be inspected by an authorised Ottobock Service Center.
Compliance
Group 1/class B
Not applicable – power
below 75 W
Product meets the require­
to
ments of the standard.
Electromagnetic environment directive
The product uses HF energy exclusively for its internal
functioning. Its HF emissions are therefore very low,
and interference with neighbouring electronic devices is
unlikely.
Appendices
79

Advertisement

Table of Contents
loading

Table of Contents