Appendices
Error while charging the product
LED on
LED on
power sup­
battery
ply
charger
Beep signal
4 x short at intervals of
approx. 20 sec. (continu­
ously)
15.2.3 Error messages while establishing a connection with the cockpit app
Error message
Component was connec­
ted
to
another
device.
Establish connection?
Mode change failed
78
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
Battery is fully charged (or connection
with product is interrupted).
Error
Charging the battery outside the allowable
temperature range
Cause
The component was connec­
ted to another mobile device
An attempt was made to
switch to a different MyMode
while the component was in
motion (e.g. while walking)
A current connection to the
prosthesis was interrupted
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check
socket
device.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Take note of the confirmation signal for
differentiation.
When the battery charger is connected or
disconnected, a self-test is conducted and
confirmed by a beep and vibration signal.
The battery is fully charged if this signal is
heard.
If no signal is emitted, the connection to
the product is interrupted.
If the connection to the product is inter­
rupted, an authorised Ottobock Service
Centre must inspect the product, battery
charger and power supply.
Resolution
Check whether the specified ambient con­
ditions for charging the battery are met
(see page 74).
Correction
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
For safety reasons, switching MyModes is only per­
mitted when components are at rest, e.g. while
standing or sitting.
Check the following points:
Distance from the prosthesis to the mobile
device
Charge level of the prosthesis
Bluetooth on the prosthesis switched on? (see
page 70)
Hold the component with the sole of the foot
facing up to make the component "visible" for
2 minutes.
Prosthesis switched on? (see page 70)
If multiple prostheses were stored, was the cor­
rect prosthesis selected?
with
another
electric
1B1-2=* Meridium

Advertisement

Table of Contents
loading

Table of Contents