6-3-3. Measure Mode Setting
Capable of setting various operating modes of the measurement. In addition, capable of setting the display mode and
comparison mode.
The measurement is controlled by the key and control inputs operations.
In addition, operation statuses can be checked with the comparator status lamps on the "Measuring" window as
follows.
Detecting Block (Detecting the peak value etc.)
key: Blinks in white.
[Detecting] lamp: Lights up in white
Holding Block (Holding the displaying value.)
key: Blinks in white.
[Detecting] lamp: Gray
(1) Operation Mode
Select the target operation mode from the following 12 types.
The WGA-910A digitally holds the peak value (maximum value) or bottom value (minimum value) of each
operation mode.
For details, see "6-3-4" and after.
<Setting Range>
Normal
Peak Hold
Block(-specified)
peak (hold)
Time(-specified)
peak (hold)
Bottom hold
Block(-specified)
bottom (hold)
Time(-specified)
bottom (hold)
Arbitrary point
hold
Block (-specified)
peak (&) bottom
(hold)
Time (-specified)
peak (&) bottom
(hold)
[Hold] lamp: Gray
[Hold] lamp: Lights up in white.
Normal measurement that measures and displays all the time.
When the HOLD command is turned ON and after the lapse of delay time, the peak
value is detected and indicated until the HOLD command is turned OFF. The detection
of the peak value and hold display are released when the hold command is turned OFF
or reset command is turned ON.
When the HOLD command is turned ON and after the lapse of delay time, the peak
value is detected until the command is turned OFF. Setting the detecting block is known
as block-specified operation. If the HOLD command is turned OFF, the indicated value
is kept hold. When the RESET command is turned ON, the kept peak value is canceled.
When the HOLD command is turned ON and after the lapse of delay time, the peak
value is detected for the preset specified time. Setting the detecting time is known as
time-specified operation. After the lapse of preset detecting time, the indicated peak
value is kept hold. When the RESET command is turned ON, the kept peak value is
canceled.
When the HOLD command is turned ON and after the lapse of delay time, the bottom
value is detected and indicated until the HOLD command is turned OFF. The detection
of the bottom value and hold display are released when the HOLD command is turned
OFF or RESET command is turned ON.
Indicates the bottom value by block-specified operation. When the HOLD command is
turned OFF, the indicated bottom value is kept hold. When the RESET command is
turned ON, the kept bottom value is canceled.
Indicates the bottom value by time-specified operation. After the lapse of preset
detecting time, the indicated bottom value is kept hold. When the RESET command is
turned ON, the indicated bottom value is canceled.
When the HOLD command is turned ON and after the lapse of delay time, the measured
value at that moment (arbitrary point) is kept hold until the HOLD command is turned
OFF.
Indicates the both bottom and peak values by the block-specified operation. When the
HOLD command is turned OFF, the indicated peak and bottom values are kept hold.
When the RESET command is turned ON, the kept hold values are canceled.
Indicates the both bottom and peak values by the time-specified operation. After the
lapse of preset detecting time, the indicated peak and bottom values are kept hold. When
the RESET command is turned ON, the indicated hold values are canceled.
65