KYOWA WGA-910A Series Instruction Manual page 74

Hide thumbs Also See for WGA-910A Series:

Advertisement

6-3-4. Descriptions of the operation mode
6-3-4-1. Normal
Normal measurement that always measures and displays value.
No control signal for the detection.
"Detect Time", "Delay Time", "Comparison Mode", "Display Mode" cannot be set.
6-3-4-2. Peak Hold
Capable of setting the "Delay time." However, the "Detect Time", "Comparison Mode", and "Display Mode" cannot be set.
(A) Behavior with control input signal
When the HOLD command is turned ON
and after the lapse of delay time, the peak
value is detected and indicated until the
HOLD command is turned OFF.
The detection of the peak value and hold
display are released when the hold command
is turned OFF or reset command is turned
ON.
MEMO
・Short circuit the terminals between the Control I/O connector No. 35 or 36 (Input COM) and No. 20 (Hold
Command). The hold command is turned ON and the WGA-910A operates with the preset measuring
mode.
・Short circuit the terminals between the Control I/O connector No. 35 or 36 (Input COM) and No. 21
(Reset). The reset command is turned ON and the detection and holding are released.
(B) Behavior with key operation
When the
time and after the lapse of the delay time, the
detecting block starts and displays the peak
value until the key is pressed for the second
time.
Press the
end the detecting block (reset).
key is pressed the first
key for the second time to
70