(4) Comparison Mode
Selects the operation block of the comparators from the two modes.
Sets the target Comparison Mode. For the "Block-specified Peak", "Block-specified Bottom", "Time-specified
Peak ", "Time-specified Bottom", "Block Averaging" and "Time Averaging" in the "Operation Mode" window
only.
For other operation modes, "Comparison Mode" setting is invalid.
<Setting Range>
Normal Comp.
Hold Comp.
<How to operate>
1) Press the
(1/2)" window.
2) Select and press the target key and press the
the setting value.
Always compares the measured value and compared value during detecting and holding block.
[Example]
The comparator output turns ON when the measured value exceeds the HI compared value,
turns OFF when the measured value lowers the compared value.
Compares the measured value and compared value during holding block.
[Example]
Even if the measured value exceeds the HI compared value during detecting block, the
comparator output does not turn ON at the time. The comparator output is turned ON when
the display value is exceeding the HI compared value and the WGA-910A is in holding
block. After resetting, the comparator output is turned OFF.