5-3. HOW TO CALIBRATE
5-3-1. Calibration Window
<How to operate>
1) Press the
2) Press the
5-3-2. Manual Input Calibration
Conducts calibration by manual setting in order of excitation voltage, no-load zero, Sensitivity registering calibration
or in order of actual load calibration and unit.
(1) Bridge Voltage (Excitation Voltage)
Capable of setting the excitation voltage that applies to the sensor from 2 V or 10 V.
Select the higher voltage within the recommended excitation voltage of the sensor.
<Setting range>
2 V or 10 V
<How to operate>
1) Press the
2) Press either
key to register the setting value.
3) After the registration, press the
"No-load Zero" window.
key on the "Measuring" window.
key on the "Setting" window.
key on the "Calibration" window.
or
key and press the
key to move to the
38
Calibration
Return
'Bridge Voltage','No-Load Zero
Manual Input Cal.
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Calibration is automatically
TEDS Auto Cal.
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
TEDS Part Cal.
the only selected item from
TEDS built-in sensor.
'TEDS Read Ope.'and 'Zero dur-
TEDS Ope.Setting
ing TEDS'are set.TEDS informa-
tions are displayed.
Next
Measure
1 / 2