(26) [SCF]
Content:
Format:
Parameter:
Return value format:
Description and others
Reading command
Return value
(27) [SCV]
Content:
Format:
Parameter:
Return value format:
Description and others
Reading command
Return value
Set the comparison setting.
SCFa, b, c, d, e, xxxxxx<CR>
a, b, c, d, e, xxxxx
ok<CR><LF> (The measuring condition selecting signal is command.)
Corresponding to the [/M] command.
PRR041
a, b, c, d, e, xxxxx
Set the compared value.
SCVa, ±xxxxxx<CR>
a, ±xxxxxx
Compared value: -99999 to 99999 (count)
Indicated value with a decimal point. If the indicated value has no
decimal point, it is filled with '0s' from the head.
Comparator selection
0: HH
ok<CR><LF>
Corresponding to the [/M] command.
PRR051, PRR052, PRR053, PRR054
a, ±xxxxxx<CR><LF>
Compared value
Comparator selection
Hysteresis width: 0 to 9999 (count)
Including a decimal point.
Comparator output logic
0: Negative logic
Using comparator LL
0: Use
1: Not use
Using comparator LO
0: Use
1: Not use
Using comparator HI
0: Use
1: Not use
Using comparator HH
0: Use
1: Not use
Hysteresis width: 0 to 9999 (count)
Including a decimal point.
Comparator output logic
0: Negative logic
Using comparator LL
0: Use
1: Not use
Using comparator LO
0: Use
1: Not use
Using comparator HI
0: Use
1: Not use
Using comparator HH
0: Use
1: Not use
1: HI
2: LO
125
1: Positive logic
1: Positive logic
3: LL