(21) [VRC]
Content:
Format:
Parameter:
Return value format:
Description and others
Reading command:
Return value
(22) [EID]
Content:
Format:
Parameter:
Return value format:
Description and others
Reading command
Return value
(23) [SSM]
Content:
Format:
Parameter:
Return value format:
Description and others
Reading command
Return value
To execute numeric value registering calibrations
VRC±x.xxxx<CR>
±x.xxxx
Initial value of transducers: -3.2000 to 3.2000 (mV/V)
ok<CR><LF>
The results of the command will be saved.
PRR011
x.xxxx<CR><LF>
Initial value of transducers:
Conduct TEDS-based automatic calibration.
EID<CR>
None
ok<CR><LF>
(when in normal)
E-13<CR><LF> (When an error occurs)
Result of this command is overwritten.
Sets the excitation voltage and sensitivity registering calibration with the rated capacity
value, rated output value, recommended excitation voltage value, and unit of the TEDS
information.
PRR010
±x.xxxx, ±yyyyyy, zz<CR><LF>
Set the smoothing function.
SSMaa, b, c<CR>
aa, b, c
Analog filter
0: 1 Hz
Minimum scale
0: 1
Moving average
00: None
04: 16 times
08 256 times
ok<CR><LF>
Corresponding to the [/M] command.
PRR020
aa, b, c<CR><LF>
Analog filter
Minimum scale
Moving average
E-00<CR><LF>(When an error occurs)
Unit
Rated display value
Rated output value
1: 30 Hz
2: 300 Hz
1: 2
2: 5
3: 10
01: 2 times
05: 32 times
09: 512times
123
3: None (1 kHz or more)
4: 20
5: 50
02: 4 times
03: 8 times
06: 64 times
07: 128 times
10: 1024 times 11: 2048 times
6: 100