(2) No-Load ZERO
Loads the output voltage of the sensor with no-load applied and registers it as a reference of a total zero. No-load
zero adjustment range is same with the measuring range.
<Setting range>
(-3.2000 to +3.2000 mV/V; display only)
<How to operate>
1) Set the sensor in no-load state.Current sensor output value is
displayed on the "New" column.
2) After stabilizing the value in the "New" column, press the
key.
The "No-Load Zero" window on the upper left blinks, the sensor
output is loaded and the zero point output at the no-load state is
registered.
3) After the registration, press either
key to move to the target calibration window.。
MEMO
The "No-Load Zero" window blinks when conducting the no-load zero with the following operations
during the measurement or during the setting.
・When conducting the no-load zero with communication.
・After setting "Execute" in "ZERO during TEDS" window and before reading the TEDS information for
conducting the TEDS-based automatic calibration or partial calibration based on TEDS calibration data.
Blinks
or
No-load Zero
Conducting no-
load Zero
Calibration is conducted with
Calibration
sensor by 'Manual Input','TEDS
Auto'or'TEDS Part'Cal.
'Smoothing','Additional Value'
Input Setting
and 'Zero Compensation'are set
The sensor output is monitored
'Comparison Set'and 'Measure
Meas.Cond.Setting
Mode Set'are set as measure-
ment condition.
The necessary setting to dis-
Wave Display Setting
play the wave such as 'X-axis'
and 'Y-axis'are set.
39
Measure
Next
1 / 2