(3) Comparator Out. Logic
Common to all comparators. Sets the output logic (open collector logic) of comparators.
<Setting Range>
Negative Logic
Positive Logic
<How to operate>
1) Press the
window.
2) Input a setting value with the numeric keyboard and press the
key for registration.
(4) Using Comparator
Select whether or not to use comparators.
If the
key is selected, the target comparator key on the "Measuring" window is displayed in gray.
<Setting Range>
Use
Compares with the measured value with the selected comparator.
Not Use
Does not compare with the measured value and does not use the selected comparator.
<How to operate>
1) Press the
window.
2)
or
target comparator.
Press the
NOTE
The comparator with selecting "Not use" in the "Using Comparator" window, compared value keys on the
"Comparison Setting" and "Measuring" windows are invalid.
For example, when selecting the HH and LL comparators to Unused, the
keys on the "Comparison Setting" window are invalid, the target keys, the current setting values, and the
descriptions are displayed in gray.
On the "Measuring" window, the
and
Comparator output is ON when the comparison result is ON. (Active low)
Comparator output is OFF when the comparison result is OFF. (Active high)
key on the "Comparison Setting (2/2)"
key on the "Comparison Setting (2/2)"
key changes by pressing the right key of the
key for registering.
keys are displayed in gray.
and
keys,
63
Comparison Setting
Return
The width that comparator is
Hysteresis Width
turned OFF the once turned ON
is set.
0.00kN
The output logic of comparator
Comparator Out.Logic
is set.
Negative Logic
Select whether each comparator
Using Comparator
is used or not.
Previous
Measure 1
and
and
lamps are in valid and
Measure
2 / 2