(3-a) Sensitivity Regist. Cal. (Sensitivity Registering Calibration)
Inputs the rated capacity and rated output described in the Test Data Sheet for equivalent calibration with the
actual load when the actual load cannot be applied to the sensor. Suppose that you are using a transducer with the
"Rated Capacity: 200 kN" and "Rated Output: 2.0040 mV/V." In order to read "200.00" against a load of 200 kN
that you will give to the transducer, do as follows.
NOTE
Set the rated display 100000 count or less per rated output 0.1 mV/V.
<Setting range>
Rated Output: -3.2000 to +3.2000 mV/V (excluding 0)
Rated Display: -99999 to 99999 (excluding 0)
<How to operate>
1)After pressing the
keyboards. In the case of example, if "2.0040" is input, "2.0040
mV/V" is displayed in the "New" column.
In addition, the
Press the
upper left blinks and the "Rated Output" is registered.
2) After pressing the
keyboards.
In the case of example, if "200.00" is input, "200.00 kN" is displayed
in the "New" column.
In addition, the
3) Press the
upper left blinks and the "Rated Output" is registered.
4) After the registration, press the
window.
MEMO
The "Sens. Registering" window blinks when conducting the sensitivity registering calibration with
communication during the measurement or during the setting.
key, input a setting value with a numeric
key blinks during the setting.
key. The "Sens. Registering" window on the
key, input a setting value with a numeric
key blinks during the setting.
key. The "Sens. Registering" window on the
key to move to the "Unit"
Sens.Regis-
Sensitivity reg-
istering
tering
Calibration is conducted with
Calibration
sensor by 'Manual Input','TEDS
Auto'or'TEDS Part'Cal.
'Smoothing','Additional Value'
Input Setting
and 'Zero Compensation'are set
The sensor output is monitored
'Comparison Set'and 'Measure
Meas.Cond.Setting
Mode Set'are set as measure-
ment condition.
The necessary setting to dis-
Wave Display Setting
play the wave such as 'X-axis'
and 'Y-axis'are set.
40
Blinks
Measure
Next
1 / 2

Advertisement

loading

This manual is also suitable for:

Wga-910a-0Wga-910a-1Wga-910a-2Wga-910a-3Wga-910a-4Wga-910a-12 ... Show all