(8) When displaying error.
During the TEDS calibration, due to the following causes, the TEDS information cannot be read, displays the
"TEDS Error" and sensitivity registering may not be conducted.
・TEDS signal lines are wrongly connected. Check the wiring by referring to "3-2 WIRING."
・The connecting number of sensors is not applicable. Check that the connecting number of sensors is within
1 to 4.
・Cable is too long.
TEDS information may not function correctly if the cable length exceeds 30 m.
Pay attention when using an extension cable to the sensor.
・Connecting non-TEDS-compatible sensor.
After disappearing the "TEDS Error" display, the calibration value of the TEDS information is not applied and
the WGA-910A operates with the calibration values that are registered before the TEDS calibration.